Chapter 5 309
Language Reference
STATus Subsystem
Operation Negative Transition
:STATus:OPERation:NTRansition <integer>
:STATus:OPERation:NTRansition?
This command determines which bits in the Operation Condition register will set
the corresponding bit in the Operation Event register when that bit has a negative
transition (1 to 0). The variable <integer> is the sum of the decimal values of the
bits that you want to enable.
Factory Preset
and *RST: 0
Range: Integer, 0 to 32767
Operation Positive Transition
:STATus:OPERation:PTRansition <integer>
:STATus:OPERation:PTRansition?
This command determines which bits in the Operation Condition register will set
the corresponding bit in the Operation Event register when that bit has a positive
transition (0 to 1). The variable <integer> is the sum of the decimal values of the
bits that you want to enable.
Factory Preset
and *RST: 32767 (all 1s)
Range: Integer, 0 to 32767
Preset the Status Byte
:STATus:PRESet
Sets bits in most of the enable and transition registers to their default state. It
presets all the Transition Filters, Enable Registers, and the Error/Event Queue
Enable. It has no effect on Event Registers, Error/Event Queue ESE, and SRE
Registers as described in IEEE Standard 488.2-1992, IEEE Standard Codes,
Formats, Protocols and Common Commands for Use with ANSI/IEEE Std
488.1-1987. New York, NY, 1992.