Chapter 1 39
Programming Fundamentals
Creating Valid Commands
Creating Valid Commands
Commands are not case sensitive and there are often many different ways of
writing a particular command. These are examples of valid commands for a given
command syntax:
Command Syntax Sample Valid Commands
[:SENSe]:BANDwidth[:RESolution]
<freq>
The following sample commands are all identical. They
will all cause the same result.
:Sense:Band:Res 1700
:BANDWIDTH:RESOLUTION 1.7e3
:sens:band 1.7KHZ
:SENS:band 1.7E3Hz
:band 1.7kHz
:bandwidth:RES 1.7e3Hz
:CALCulate:MARKer[1]|2|3|4:Y? The last command below returns different results than
the commands above it. The number 3 in the command
causes this. See the command description for more
information.
:CALC:MARK:Y?
:calc:mark:y?
:CALC:MARK2:Y?
[:SENSe]:DETector[:FUNCtion]
NEGative|POSitive|SAMPle
DET:FUNC NEG
:Sense:Detector:Function Sample
:INITiate:CONTinuous OFF|ON|0|1 The sample commands below are identical.
:INIT:CONT ON
:init:continuous 1

Table of Contents

Other manuals for Agilent Technologies E7401A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E7401A and is the answer not in the manual?

Agilent Technologies E7401A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE7401A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals