EasyManua.ls Logo

Agilent Technologies E7401A

Agilent Technologies E7401A
349 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 1 39
Programming Fundamentals
Creating Valid Commands
Creating Valid Commands
Commands are not case sensitive and there are often many different ways of
writing a particular command. These are examples of valid commands for a given
command syntax:
Command Syntax Sample Valid Commands
[:SENSe]:BANDwidth[:RESolution]
<freq>
The following sample commands are all identical. They
will all cause the same result.
:Sense:Band:Res 1700
:BANDWIDTH:RESOLUTION 1.7e3
:sens:band 1.7KHZ
:SENS:band 1.7E3Hz
:band 1.7kHz
:bandwidth:RES 1.7e3Hz
:CALCulate:MARKer[1]|2|3|4:Y? The last command below returns different results than
the commands above it. The number 3 in the command
causes this. See the command description for more
information.
:CALC:MARK:Y?
:calc:mark:y?
:CALC:MARK2:Y?
[:SENSe]:DETector[:FUNCtion]
NEGative|POSitive|SAMPle
DET:FUNC NEG
:Sense:Detector:Function Sample
:INITiate:CONTinuous OFF|ON|0|1 The sample commands below are identical.
:INIT:CONT ON
:init:continuous 1

Table of Contents

Other manuals for Agilent Technologies E7401A

Related product manuals