Chapter 5 261
Language Reference
MMEMory Subsystem
Front Panel
Access:
File, Copy
Move Data to File
:MMEMory:DATA <file_name>,<definite_length_block>
:MMEMory:DATA? <file_name>
Loads <definite_length_block> into the memory location <file_name>.
The query returns the contents of the <file_name> in the format of a definite length
block. This command can be used for copying files out of the analyzer over the
remote bus. Refer to chapter 3, Programming Examples, for more information.
Example: Load abcd into C:source.txt:
:MMEM:DATA C:source.txt,#14abcd
Delete a File
:MMEMory:DELete <file_name>
Delete a file.
Example: :MMEM:DEL C:source.txt
Remarks: If <file_name> does not exist, a File Name Error will occur.
Front Panel
Access:
File, Delete
Load a Corrections Table from a File
:MMEMory:LOAD:CORRection
ANTenna|CABLe|OTHer|USER,<file_name>
Loads the data in the file <file_name> to the specified correction set.
Example: :MMEM:LOAD:CORR ANT, A:TEST5.CBL
Front Panel
Access:
File, Load, Type, Corrections
Load a Limit Line from Memory to the Instrument
:MMEMory:LOAD:LIMit LLINE1|LLINE2,<file_name>
Loads a limit line, from the specified file in mass storage to the instrument.
Loading a time limit line deletes any frequency limit lines. Similarly, loading a
frequency limit line deletes any time limit lines.
Example:. :MMEM:LOAD:LIM LLINE2,C:mylimit.lim
Remarks:. There is no SCPI short form for parameters LLINE1|LLINE2.
Front Panel
Access:.
File, Load, Type, Limits