Chapter 5 337
Language Reference
TRIGger Subsystem
Front Panel
Access:.
Trig, Video
Video Trigger Level Frequency
:TRIGger[:SEQuence]:VIDeo:LEVel:FREQuency <freq>
:TRIGger[:SEQuence]:VIDeo:LEVel:FREQuency?
This command is used to adjust the Video Trigger level when in FM Demod, and
Demod View is on.
Default Unit: Hz
Remarks: Video is adjusted using this command, but must also be selected
using the command :TRIGger[:SEQuence]:SOURce
VIDeo. When not in FM Demod, the Video Trigger level is
adjusted/queried using the command
:TRIGger[:SEQuence]:VIDeo:LEVel <ampl>.
NOTE Trigger Delay is not available in Video trigger mode, so turning Video on turns off
Trigger Delay, but preserves the value of Trigger Delay.