EasyManuals Logo

Agilent Technologies E7401A Programmer's Guide

Agilent Technologies E7401A
349 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #279 background imageLoading...
Page #279 background image
Chapter 5 279
Language Reference
[:SENSe]:CORRection Subsection
Input Impedance Correction
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude] <number>
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude]?
Amplitude correction is applied to the display data to adjust for measurement
situations where the unit under test has a different impedance than the 50 input
impedance of the analyzer.
Factory Preset
and *RST: The factory default is the input impedance of the analyzer.
Range: 50 or 75 ohms
Default Unit: ohms
Front Panel
Access:
Input, Input Z Corr 50 75
External Amplifier Correction
[:SENSe]:CORRection:OFFSet[:MAGNitude] <rel_ampl>
[:SENSe]:CORRection:OFFSet[:MAGNitude]?
A single value of amplitude correction can be applied to the displayed trace data to
compensate for signal losses or gains that are due to other devices in the
measurement setup, rather than the unit under test.
Factory Preset
and *RST: 0 dB
Range: 81.9 to 81.9
Default Unit: dB
Front Panel
Access:
AMPLITUDE/Y Scale, Ext Amp Gain

Table of Contents

Other manuals for Agilent Technologies E7401A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E7401A and is the answer not in the manual?

Agilent Technologies E7401A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE7401A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals