[:SENSe]:FREQuency Subsection
Stop Frequency
[:SENSe]:FREQuency:STOP <freq>
[:SENSe]:FREQuency:STOP?
Set the stop frequency.
Factory Preset
and *RST: 1.06 GHz
Range: EMC E7401A: 80 MHz
1
to 1.58 GHz
EMC E7402A: 80 MHz
1
to 3.10 GHz
EMC E7403A: 80 MHz
1
to 6.78 GHz
EMC E7404A: 80 MHz
1
to 13.3 GHz
EMC E7405A: 80 MHz
1
to 27.0 GHz
Default Unit: Hz
Front Panel
Access:
FREQUENCY/Channel, Stop Freq
Frequency Synthesis Mode
[:SENSe]:FREQuency:SYNThesis 1|2|3
[:SENSe]:FREQuency:SYNThesis?
This command switches between two phase noise optimization modes. Mode 2
optimizes the analyzer for close-in phase noise. Mode 3 optimizes the analyzer for
tuning speed. Mode 1 is not recommended for new designs.
This command is available for the following models only:
E7402A, E7403A, E7404A, E7405A
Factory Preset
and *RST: 3
History: Added with firmware revision A.08.00.
Front Panel
Access:
AUTO COUPLE, PhNoise Opt
1. 10 Hz minimum in log sweep mode.

Table of Contents

Other manuals for Agilent Technologies E7401A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E7401A and is the answer not in the manual?

Agilent Technologies E7401A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE7401A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals