Agilent 1260 Infinity VWD User Manual 61
Using the Detector
4
Setting up an Analysis
Figure 19 Detector Settings (default)
1 signal with individual wavelength setting
stop and post time can be set (if required)
peakwidth depends on the peaks in the
chromatogram, see “Peakwidth Settings” on
page 71.
time table for programmable actions during the run
Zero Offset Limits: 1 – 99 % in steps of 1 %
Attenuation Limits: 0.98 – 4000 mAU at discrete
values for either 100 mV or 1 V full scale
additional signals can be stored with the normal
signal (for diagnostics)
autobalance to zero absorbance (on the analog
output plus offset) at begin and/or end of run
see “Special Setpoints” on page 70.