62 Agilent 1260 Infinity VWD User Manual
4 Using the Detector
Setting up an Analysis
11 Pump the water/acetonitrile (30/70 %) mobile phase through the column
for 10 minutes for equilibration.
12 Click the button and select Change... to open the Signal Plot
information. Select the Pump: Pressure and the VWD A: Signal 254 as
signals. Change the Y- range for the VWD to 1 mAU and the offset to 20
% and the pressure offset to 50 %. The X- axis range should be 15
minutes. Press OK to exit this screen.
Figure 20 Edit Signal Plot Window