EasyManuals Logo
Home>Agilent Technologies>Portable Generator>N5182A

Agilent Technologies N5182A User Manual

Agilent Technologies N5182A
175 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #105 background imageLoading...
Page #105 background image
94 Agilent N5181A/82A MXG Signal Generators User’s Guide
Basic Digital Operation (Option 651/652/654)
Using Waveform Markers
Setting Marker Polarity
Setting a negative marker polarity inverts the marker signal.
1. In second Arb menu (page 87), press Marker Polarity.
2. For each marker, set the marker polarity as desired.
The default marker polarity is positive.
Each marker polarity is set independently.
See also, “Saving Marker Polarity and Routing Settings” on page 83.
As shown on page 93:
Positive Polarity: On marker points are high (3.3V).
Negative Polarity: On marker points are low (0V).
RF blanking always occurs on the low part of the signal regardless of the polarity setting.

Table of Contents

Other manuals for Agilent Technologies N5182A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5182A and is the answer not in the manual?

Agilent Technologies N5182A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5182A
CategoryPortable Generator
LanguageEnglish

Related product manuals