Agilent N5181A/82A MXG Signal Generators User’s Guide 147
Troubleshooting
Internal Media Data Storage
Internal Media Data Storage
Instrument State Saved but the Register is Empty or Contains the Wrong State
If the register number you intended to use is empty or contains the wrong instrument state, recall
register 99. If you selected a register number greater than 99, the signal generator automatically saves
the instrument state in register 99.
See also, “Working with Instrument State Files” on page 38.
External Media Data Storage
Instrument Recognizes External Media Connection, but Does Not Display Files
If the external media works on other instruments or computers, it may simply be incompatible with
the signal generator; use a different memory stick. See the signal generator data sheet for information
on compatible media.
Preset
The Signal Generator Does Not Respond
If the signal generator does not respond to a preset, the instrument may be in remote mode, which
locks the keypad.
To exit remote mode and unlock the preset keys, press Local Cancel/(Esc).
Pressing Preset Performs a User Preset
This behavior results from the use of a backward- compatible SCPI command. To return the signal
generator to normal use, send the command :SYST:PRESet:TYPE NORM.
For information on SG−1364/U SCPI commands, refer to the SCPI Command Reference.