EasyManuals Logo

Ametek JOFRA RTC-156 B User Manual

Ametek JOFRA RTC-156 B
173 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #102 background imageLoading...
Page #102 background image
127139 13 2014-07-30 101
The following guidelines must be observed before emptying the well :
1. Switch off the calibrator using the power control switch.
2. Before handling the fluid, it must be cooled down to a
temperature close to ambient.
3. Remove the sensor basket and clean it with disposable paper
towels.
4. Remove the stirring magnet using the stirring magnet remover
supplied and clean it with disposable paper towels.
5. Empty the well using the liquid drainage tube supplied. Tilting
the calibrator is not recommendable, as it increases the risk of
splashing oil all over the test area.
Caution…
Avoid getting silicone oil on the clothes. It is impossible to
wash off.
6. Any remaining oil in the well is cleaned up using disposable
paper towels. It is recommendable to use the optional
cleaning oil when cleaning the well

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek JOFRA RTC-156 B and is the answer not in the manual?

Ametek JOFRA RTC-156 B Specifications

General IconGeneral
BrandAmetek
ModelJOFRA RTC-156 B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals