The following guidelines must be observed before emptying the well :
1. Switch off the calibrator using the power control switch.
2. Before handling the fluid, it must be cooled down to a
temperature close to ambient.
3. Remove the sensor basket and clean it with disposable paper
towels.
4. Remove the stirring magnet using the stirring magnet remover
supplied and clean it with disposable paper towels.
5. Empty the well using the liquid drainage tube supplied. Tilting
the calibrator is not recommendable, as it increases the risk of
splashing oil all over the test area.
Caution…
Avoid getting silicone oil on the clothes. It is impossible to
wash off.
6. Any remaining oil in the well is cleaned up using disposable
paper towels. It is recommendable to use the optional
cleaning oil when cleaning the well