127139 13 2014-07-30 89
4.11.4 Setting calibration interval
Sets the required recalibration interval for the calibrator.
Choose a value between 1 month and 99 months.
When the recalibration interval is exceeded, the warning
symbol will appear in the display.
Note…
The recalibration interval is not used for the external
reference sensor and the DLC. The interval for these
sensors are stored in the intelligent sensor.
4.11.5 Changing the date and time
Date:
Use the NUMERIC keys to enter a new date.
The date can only be entered using the format yyyy-mm-dd.
When entering the date with different format, the text will
disappear when you try to accept the setting.
Time:
The calibrator is set up with a default time (present time).
Use the NUMERIC keys to enter a new time using the
format hh.mm.
Time Zone:
The relevant time zone is selected from a list of various
zones.