34 2014-07-30 127139 13
Tables of recommended oil level @23°C well temperature
RTC-250 A/B/C RTC-158 A/B/C
For recommended
50 cSt oil
For recommended
10 cSt oil
0°C - 50°C
100%
-20°C - 50°C 100%
50°C - 100°C 95%
50°C - 100°C 95%
100°C - 150°C 90%
100°C - 120°C 90%
150°C - 200°C 85%
120°C - 155°C 85%
200°C - 250°C 80%
For the best result, the oil should be in good condition and free of
foreign objects. If water is accumulated in the oil due to melted ice
crystals – the oil must be dried out, by heating it up.
Warning
Do not handle hot fluid.
If the fluid is heated beyond the flash point, it may
constitute a fire hazard.
If the fluid has caught fire, switch off the main power to prevent further
heating of the fluid. Flames are best extinguished by cowering the well
with a non-flammable lid.
Warning
Do not under any circumstances pour water on burning oil. It
might cause a dangerous steam explosion.
3.3.3 Handling of lids
It is strongly recommended to leave the lid on during calibration.
Calibration without the lid may affect the temperature stability and
homogeneity.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek JOFRA RTC-156 B and is the answer not in the manual?

Ametek JOFRA RTC-156 B Specifications

General IconGeneral
BrandAmetek
ModelJOFRA RTC-156 B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals