4.9 Auto step menu
Auto step is used to step automatically between a range of different
calibration temperatures. This is useful when calibrating sensors in
places that are difficult to reach and sensors where the output is
displayed in a different location.
Fig. 13
4.9.1 Running an Auto step calibration
Access the Auto Step Setup menu by selecting “Autostep
4
from the main menu.
The Auto Step Setup menu is displayed.
Temperature stable
Calibrator starts working towards next temperature
t
T
T
T
T
T
T [°C/°F]
Time
[min]
Start
1
2
3
4