127139 13 2014-07-30 103
Replace the fuses.
RTC-156/157/158/159: 115V 8AT = 127211 / 230V, 4AT = 127210
RTC-250/700: 115V, 10AF = 60B302 / 230V, 5AF = 127573
If the fuses blow immediately after you have replaced them,
the calibrator should be returned to the manufacturer for
service.
6.1 Error messages
Error messages are displayed in a dialog box with the following text:
Internal Error # xxx
Please read the manual for further information
Error # Error text Category Solution
0
Read temperature lower
than calibrator minimum
temperature.
Warning Please report the error to your
local distributor or to AMETEK
Denmark’s service department.
1
Read temperature
higher than calibrator
maximum temperature.
Warning Please report the error to your
local distributor or to AMETEK
Denmark’s service department.
2
Read temperature
higher than current SET-
temperature.
Warning Please report the error to your
local distributor or to AMETEK
Denmark’s service department.
3
Zone 1 or zone 2
temperature deviation.
Warning The calibrator could be stressed
due to the insertion of too many
sensors. Remove some of the
sensors. If the error still occurs
please report the error to your
local distributor or to AMETEK
Denmark’s service department.
4
Internal reference
measuring circuit error.
Warning Please report the error to your
local distributor or to AMETEK
Denmark’s service department.
5
Internal reference
sensor error.
Warning Please report the error to your
local distributor or to AMETEK
Denmark’s service department.