EasyManua.ls Logo

Ametek JOFRA RTC-156 B

Ametek JOFRA RTC-156 B
173 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
78 2014-07-30 127139 13
4.10.2 Setting the parameters for TRUE – reference sensor (B
and C versions only)
Sensor type:
Internal reference source.
The internal reference sensor will be displayed as the TRUE
value on the main screen.
The calibrator has a set of internal stability criteria it shall
meet before stability is indicated.
Additional stability time
may be set beyond the internal stability criteria.
Set the additional stability time by pressing and the
NUMERIC keys. Stability time can be set (in minutes) using
integers from 0 – 99.
External reference source
The TRUE value on the main screen will be read from the
Intelligent Reference Sensor connected to the REF. INPUT
on the front panel (see section 4.2 fig. 6). The calibrator
automatically reads the calibration data and serial number of
the Sensor.
Convert to temperature:
“yes” sets the readout of the External reference as a
temperature.
“no” sets the readout of the External reference in
values.

Table of Contents

Related product manuals