Caution…
If the speed level chosen is too high, the magnet will fall off
making a rattling sound and there will be no stirring in the
fluid. With no stirring of the fluid, temperature gradients will
emerge in the bath, which will again affect the result of the
calibration.
To reconnect the magnet, set the speed level to 0 and
select a speed setting lower than the previous.
Press or to accept the value and return to the
setup menu.
4.13 Information Screen
Information about the status can be viewed using the Info
function from the vertical menu (press ).
A status summary of the sensors setting and stability
information is displayed.
If a warning or an error has occurred, it will be listed on the
information screen.
Note…
The list will be cleared, when the calibrator is switched off.