127139 13 2014-07-30 105
Denmark’s service department
Error # Error text Category Solution
102
Reference input board
error.
Error Please report the error to your
local distributor or to AMETEK
Denmark’s service department.
103
The reference input
board has not been
calibrated
Error The reference input needs to be
calibrated.
Please report the error to your
local distributor or to AMETEK
Denmark’s service department
104
The calibration for the
heat source has expired.
Warning Calibrate the heat source
105
The calibration for the
sensor input board has
expired.
Warning Calibrate the sensor input
(mA/Ω/mV/V)
106
The calibration for the
reference input board
has expired.
Warning Calibrate the reference input
107
The calibration for the
external reference
sensor has expired.
Warning Calibrate the external reference
sensor
108
The calibration for the
DLC has expired.
Warning The instrument needs to be
calibrated
109
Heat source not
calibrated
Error The instrument needs to be
recalibrated.
6.2 Returning the calibrator for service
When returning the calibrator to the manufacturer for service, please
enclose a fully completed service information form. Simply copy the
form on the following page and fill in the required information.
The calibrator should be returned in the original packing.
The RTC-159 contains the flammable refrigerating gas R-1270 and the
gas R-704. The amount of gas is less than 100g and it is considered
not subject to the Dangerous Goods Regulations. However this must
be declared when shipping.