88 2014-07-30 127139 13
SENSOR visible:
Choose between :
Visible
Hidden
If the Hidden option is chosen the Sensor Under Test
reading will not be displayed on the main screen.
4.11.3 Setting the sound, volume and operating mode
Sound:
Choose between :
On
Off
Enables the calibrator to make a sound during operation.
Sound responses are given at the following conditions:
Stability
Warnings
Accept of data entry
Reject of data entry
Volume:
The volume of the sound can be adjusted from 0 – 100%.
Operating mode:
Choose between :
Fast
Silent
“Fast” – the fan operates in a fast mode giving the best
performance of cooling.
“Silent” – the fan operates in a silent mode reducing the
noise. Using this option the cooling process is
made a little slower and the calibrator might not
be able to reach the specified minimum
temperature.