INPUT SPECIFICATIONS
ACCURACY IN C/F
Accuracy RTD
Pt100 MILL: -200°C to 850°C
(excluding sensor accuracy)
±0.020°C(±0.036°F) @ -200°C(-328°F)
±0.023°C(±0.041°F) @ -90°C(-130°F)
±0.024°C(±0.043°F) @ -50°C(-58°F)
±0.026°C(±0.046°F) @ 0°C(32°F)
±0.030°C(±0.054°F) @ 155°C(311°F)
±0.035°C(±0.063°F) @ 320°C(608°F)
±0.038°C(±0.068°F) @ 420°C(788°F)
±0.046°C(±0.083°F) @ 660°C(1220°F)
±0.048°C(±0.086°F) @ 700°C(1292°F)
±0.054°C(±0.097°F) @ 850°C(1562°F)
Accuracy RTD YSI-400
15°C to 50°C
(excluding sensor accuracy)
±0.002°C(±0.004°F) @ 15°C(59°F)
±0.008°C(±0.013°F) @ 50°C(122°F)
Accuracy Pt100 reference input
(excluding sensor accuracy)
±0.006°C(±0.010°F) @ -200°C(-328°F)
±0.007°C(±0.013°F) @ -90°C(-130°F)
±0.008°C(±0.015°F) @ -50°C(-58°F)
±0.008°C(±0.015°F) @ 0°C(32°F)
±0.011°C(±0.019°F) @ 155°C(311°F)
±0.014°C(±0.024°F) @ 320°C(608°F)
±0.015°C(±0.027°F) @ 420°C(788°F)
±0.020°C(±0.035°F) @ 660°C(1220°F)
±0.020°C(±0.036°F) @ 700°C(1292°F)
±0.023°C(±0.042°F) @ 850°C(1562°F)
* Available upon request on selected markets.