Start the stirring magnet by following the procedure in section
4.12.
Place the calibration lid onto the well. See section 3.3.3 for
drilling information.
Warning
To avoid hazards from treating fluids in a wrong manner,
always reduce the "Max. SET-temperature allowed” in the
CALIBRATOR SETUP MENU according to the specifications
of the fluid to be used.
If using a calibrator outside of the fluids specifications there
is a risk of fire hazards, personal Injury or chemical release.
By reducing the "Max. SET-temperature allowed”, the
calibrator cannot be used outside this temperature range.
Be aware of the flash point, the boilling point and other fluid
properties applicable to the usage when setting the Max.
SET-temperature. Read the MSDS (Material Safety Data
Sheet) of the liquid before use.
Select a SET-temperature according to the tables of
recommended oil volume by following the procedure in
section 4.6.
Warning
Ensure that the sensor is absolutely clean and dry, as a few
drops of water might cause a steam explosion.
Place the sensor to be calibrated vertically into the well. It is
recommended to use the optional support rod set for a
correct position during calibration. See fig. 5.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek JOFRA RTC-156 B and is the answer not in the manual?

Ametek JOFRA RTC-156 B Specifications

General IconGeneral
BrandAmetek
ModelJOFRA RTC-156 B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals