EasyManua.ls Logo

Ametek OXYvisor - BOS Sensor Technical Performance Specifications (Determines Operating Range for Oxyvisor)

Ametek OXYvisor
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
Installation, Operation & Maintenance Manual
OXYvisor Optical Oxygen Analyzer
Barben Oxygen Sensors (BOS), are sold separately or as parts of an integrated (SCP) package with the
OXYvisor. The sensors consist of a ber optic cable with SMA termination at one end, for connection to the
OXYvisor, and the other end, integrated with a oxygen sensing luminophore to be placed into the process or
sample stream. There are three oxygen sensors, each with unique measurement ranges that can be used with
the OXYvisor; BOS1, BOS2 and BOS3. Their selection and pairing with the OXYvisor will dene the range,
accuracy, and repeatability of the OXYvisor. For additional information on BOS sensors please refer to the
BOS sensor product data sheet.
BOS1 Sensor Specications - Liquid Phase / Gas Phase
Dissolved Oxygen (DO) Gas Phase @ 1atm, 20ºC
Measurement Range 0 - 2.0 mg/L (ppm)
0 - 5.0% O
2
(0 - 50.7 hPa)
Limit of Detection 1.0 µg/L (ppb)
0.002 % O
2
(0.02 hPa)
Resolution @ 20ºC and 1013 hPa
± 0.30 at 1 μg/L (ppb)
± 0.63 at 200 μg/L (ppb)
± 0.0007 % O
2
at 0.002 % O
2 ,
± 0.0015 % O
2
at 0.02 % O
2 ,
± 0.007 hPa at 0.023 hPa,
± 0.015 hPa at 2.0 hPa
Response Time (T
90
) < 30 sec. < 6 sec.
Accuracy @ 20ºC 1 ppb (l), 0.002 % O
2
(g),
or 3% of the measured value whichever is greater
Drift from Photo-decomposition < 1.0 ppb within 30 days (1 min sample rate)
Operating Temperature Range 0 to 50ºC (32 to 122ºF)
Allowable Sensor Temperature 90ºC (194ºF) non-continuous
BOS2 Sensor Specications - Liquid Phase / Gas Phase
Dissolved Oxygen Gaseous & Dissolved Oxygen @ 1atm, 20ºC
Measurement Range 0 - 45 mg/L (ppm) 0 - 100 % O
2
(0 - 1013 hPa)
Limit of Detection (LOD) 15 ppb dissolved oxygen 0.03 % O
2
Resolution @ 20ºC and 1013 hPa
± 4.5 at 90 μg/L (ppb)
± 0.15 at 23 mg/L (ppm)
± 0.01 % O
2
at 0.21 % O
2
± 0.1 hPa at 2 hPa
± 0.1 % O
2
at 20.9 % O
2
± 1 hPa at 207 hPa
Response Time (T
90
) < 30 sec. < 6 sec.
Accuracy @ 20ºC ± 0.4 % O
2
at 20.9 % O
2
, ± 0.05 % O
2
at 0.2 % O
2
Drift from Photo-decomposition < 0.03 % O
2
within 30 days (1 min sample rate)
Operating Temperature Range 0 to 50ºC (32 to 122ºF)
Allowable Sensor Temperature 90ºC (194ºF) non-continuous
BOS3 Sensor Specications - Liquid Phase / Gas Phase
Gas Phase Oxygen @ 1atm, 20ºC
Measurement Range 0 - 300 ppm with over-range of 1000 ppm
Limit of Detection (LOD) 0.5 ppm O
2
Resolution @ 20ºC & 1013 hPa 10 ± 0.5 ppm; 100 ± 0.8 ppm; 200 ± 1.5 ppm
Response Time (T
90
) < 3 sec. based on 0 - 300 ppm measurement range
Accuracy @ 20ºC,1 atm ± 2ppm or ± 5% of measured value whichever is greater (or as partial pressure, +/- 0.002 hPa)
Drift from Photo-decomposition < 1.5 ppm within 30 days (1 min sample rate)
Operating Temperature Range 0 to 50ºC (32 to 122ºF)
Allowable Sensor Temperature 90ºC (194ºF) non-continuous
Cross Sensitivity for BOS1, BOS2, BOS3 Sensors Listed above
No cross-sensitivity for carbon dioxide (CO
2
), hydrogen sulde (H
2
S), ammonia (NH
3
), gaseous sulfur dioxide (SO
2
), no
cross-sensitivity to pH (1-14), ionic species like sulde, sulfate, or chloride. Usable in methanol, ethanol-water mixtures,
and in pure methanol & ethanol. Avoid organic solvents like benzene, chloroform, toluene, acetone, and methylene
chloride along with any strong oxidizers such as gaseous chlorine (Cl
2
).
2.2 BOS Sensor Technical Performance Specicaons (determines operang range for OXYvisor)
󰀛
Section 2
Technical Product Specication

Table of Contents

Related product manuals