81
Installation, Operation & Maintenance Manual
OXYvisor Optical Oxygen Analyzer
To read Process Pressure, Process Temperature and any error codes present
To turn ON Relay
#1, Write the value
1 to register 44010.
This will select Relay
number as 1.
Now select the Test
position of the Relay
using register 44011.
Usually the Relays are
“Normally Open”. So
to activate Relay #1,
we need to close it.
The Value to Write is
1 to close. (0 for open
in case the relay is
“normally closed”)
A7.5.2 Turning On Relays