19
Installation, Operation & Maintenance Manual
OXYvisor Optical Oxygen Analyzer
2.8 Product Label / Name Plate - (Pending Final Approvals)
Class I, Division 2, Groups A, B, C, & D T4A
2.8.1 - Product Label / Name Plate Overview (Pending Final Approvals)
2�7�3 Class I, Division 2 Groups A,B,C, & D
IECEx Zone 2, Group IIC Markings:
The Product Label (Name Plate) is found at the top of the analyzer (Figure 4). The product label identies
important information and attributes of the specic model and should be reviewed in detail prior to installation.
US Standards: UL 12.12.01 - Edition 9 Issued 2017/09/15
CAN Standards: CSA C22.2 No. 213-17 - Edition 3 Issued 2017/09/15
The OXYvisor is certied as Process Control Equipment for use in hazardous locations (QUZW, QUZW7)
Class I, Division 2 Groups A, B, C, & D
OXYvisor
II 2 G Ex db op is IIC T4 Gb
Ex db op is IIC T4 Gb
Ta = -20°C to +55°C, IP 66, NEMA 4x
Class I Zone 1 AEx db op is IIC T4 Gb
Ex db op is IIC T4 Gb
Installation DWG# 2P0345
0539
IECEx UL 18.0033
DEMKO 18 ATEX 2031
AVERTISSEMENT:
Ne pas ouvrir quand une atmosphère
explosive est présente. Sceller toutes les sorties de câbles
conformément aux codes gouvernementaux et locaux.
WARNING: Do not open when an explosive
atmosphere is present. Seal all cable exits in
accordance with government and local code.
Made in Carson City NV - USA
www.BarbenAnalytical.com
Model#
Serial #
Power:
RoHS
WARNING: Consult Conditions of Safe Use in installation
manual and DWG# 2P0345 to preserve the protection
aorded by the equipment.
AVERTISSEMENT: Consultez les conditions d'utilisation
sécurisée du manuel d'installation et le DWG n° 2P0345
pour préserver la protection oerte par l'équipement.
BOA-DC-1-SFP-B-SI
YYMM -0000-1234XYZ
21.6 - 26.4 VDC
5 Watts
E502700 Process Control Equipment for
Use in Zone Classied Hazardous Locations
Branding Logo
Parent Company
Logo
Product Name
Serial # contains the
date of manufac-
ture. The rst four
numbers are YYMM
Power: Acceptable
range, power type
and max wattage
drawn
Installation Drawing
contains critical informa-
tion about the installation
requirements. This must
be reviewed prio to instal-
laltion.
Important Warnings! that
should be adhered to at
all times.
Hazardous Area Markings
Figure 6 - OXYvisor Product Label / Name Plate
Section 2
Technical Product Specication
Ex ec ic op is IIC T4 Gc
Class 1 Zone 2 AEx ec ic op is IIC T4 Gc
Class 1 Zone 2 Ex ec ic op is IIC T4 Gc
Certicate Number: IECEx UL 18.003Xx
North American Zone 2, Group IIC Markings:
Standards: IEC 60079-0, Edition 6, Issued 2014-10
IEC 60079-7, Edition 5.1, Issued 2014-06
IEC 60079-11, Edition 6, Issued 2014-10
IEC 60079-28, Edition 2, Issued 2015-05
IEC 61010-2-201 Edition 1, Issued 2013
US Standards: UL 60079-0, Edition 6, Dated 2017/10/20
UL 60079-7, Edition 5.1, Dated 2017/04/21
UL 60079-11, Edition 6, Dated 2014/03/28
UL 60079-28, Edition 2, Dated 2017/12/08
CAN Standards: CSA C22.2 No. 60079-0, Edition 3, Issue Date 2015/10/01
UL 60079-7 Edition 2, Issue Date 2016/
UL 60079-11, Edition 2, Issue Date 2014/02/01
UL 60079-28, Edition 1, Issue Date 2016/