A3 - Dri and Sample Rate
Sample Rate - [Menu Structure: Measurement/Sampling Rate]
The sample rate determines how often the measurement will be taken. The rate can be set to take a
measurement every 3 seconds to once every hour. It is benecial to set the sample rate as slow as possible
to minimize measurement drift between calibrations and extending the life of the sensor. Measurement
drift and sensor life are directly a󰀨ected by the sample rate due to the photo-bleaching of the measurement
element. Photo-bleaching occurs every time the blue LED light ashes to take a measurement. The factory
recommends a sample rate of every 60 seconds unless the application requires a quicker measurement. A
table is shown here as an example of estimated drift between calibrations as a function of sample rate. Please
note, that this is just an estimation based on common operating conditions.
󰀚󰀟