Installation, Operation & Maintenance Manual
OXYvisor Optical Oxygen Analyzer
Section 3
Quick Start Guide
10. The next screen will allow you to Enter Span Gas O2 Concentration. Enter the 2’nd Cal Point Span
Gas value here.
11. Start owing the Span Gas and press the enter button to Input Span Gas and to see live values.
Monitor phase angle and the temperature readings until stable and press the Enter Key to store these
values.
12. The stability of the phase angle should be +/- 0.07 and Temperature should be +/-0.2 C
13. Press the escape key to return tho the measurement display screen. If the Span Gas is still owing,
you should see this oxygen value on the measurements.
14. The sample rate will be automatically updated to the value set in Step # 4 of the Setup procedure.
Regular Maintenance:
Do NOT attempt to clean the sensing tip with any hard or abrasive object that may damage the
sensor.
If coated with oil or particulates, rinse carefully and daub with clean, wet (water or isopropyl alcohol)
towel.