Chapter 4 — Cable & Antenna Analyzer Mode Commands
S331L PM PN: 10580-00322 Rev. F 4-15
:CALCulate:TRANsform:DISTance:CABLoss <cable-loss>{
dB/m,dB/ft}
:CALCulate:TRANsform:DISTance:CABLoss?
Cmd Description: Command sets the cable loss. Verify that the appropriate unit is used.
Query Description: Query returns the cable loss based on the current units.
Param Type: Float
Default Value: 0
Default Units: dB/m
Example: To set cable loss to 2:
:CALC:TRAN:DIST:CABL 2
:CALCulate:TRANsform:DISTance:DMAX?
Query Description: Query returns max distance in the current units.
Example: To query max distance:
:CALC:TRAN:DIST:DMAX?
:CALCulate:TRANsform:DISTance:PVELocity
<propagation-velocity>
:CALCulate:TRANsform:DISTance:PVELocity?
Cmd Description: Command sets the propagation velocity.
Query Description: Query returns the propagation velocity.
Param Type: Float
Default Value: 1
Example: To set propagation velocity to 0.05:
:CALC:TRAN:DIST:PVEL 0.05
:CALCulate:TRANsform:DISTance:RESolution?
Query Description: Query returns distance resolution in the current units for reflection.
Example: To query distance resolution:
:CALC:TRAN:DIST:RESolution?

Table of Contents

Other manuals for Anritsu Site Master S331L

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Site Master S331L and is the answer not in the manual?

Anritsu Site Master S331L Specifications

General IconGeneral
BrandAnritsu
ModelSite Master S331L
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals