Chapter 4 — Cable & Antenna Analyzer Mode Commands
S331L PM PN: 10580-00322 Rev. F 4-19
:DISPlay:WINDow:TRACe:STATe TRACe|MEMory|BOTH
:DISPlay:WINDow:TRACe:STATe?
Cmd Description: Command sets the trace display type for the active trace. TRAC - only
active trace is shown; MEM - only memory trace is shown; BOTH - both
active and memory traces are shown. Use
:CALCulate:MATH:MEMorize before changing the display type.
Query Description: Query returns the trace display type.
Param Range: TRACe|MEMory|BOTH
Default Value: TRACe
Example: To show both active and memory trace on the display:
:DISP:WIND:TRAC:STAT BOTH
:DISPlay:WINDow:TRACe:Y[:SCALe]:AUToscale
Cmd Description: This command sets the top and bottom values such that the trace
occupies a significant portion of the grid.
Example: To perform auto-scale:
:DISP:WIND:TRAC:Y:AUT
:DISPlay:WINDow:TRACe:Y[:SCALe]:BOTTom <amplitude>{dB}
:DISPlay:WINDow:TRACe:Y[:SCALe]:BOTTom?
Cmd Description: Command sets the display bottom value.
Query Description: Query returns the display bottom value.
Param Type: Float
Default Value: Return Loss = 60dB; DTF Return Loss = 60dB; Cable Loss = 30dB;
VSWR = 1; DTF VSWR = 1; Transmission = -90dB; Phase=-225
Default Units: dB
Example: To set trace bottom value in Return Loss to 40 dB:
:DISP:WIND:TRAC:Y:BOTT 40
:DISPlay:WINDow:TRACe:Y[:SCALe]:FULLscale
Cmd Description: Sets the top and bottom to their maximum values.
Example: To perform full-scale:
:DISP:WIND:TRAC:Y:FULL