EasyManuals Logo

Anritsu Site Master S331L Scpi Programming Manual

Anritsu Site Master S331L
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
Chapter 4 — Cable & Antenna Analyzer Mode Commands
4-26 PN: 10580-00322 Rev. F S331L PM
[:SENSe]:CORRection:TYPE STANdard|FLEX
[:SENSe]:CORRection:TYPE?
Cmd Description: Command sets correction type.
Query Description: Query returns correction type. Query returns STAN|FLEX.
Param Range: STANdard|FLEX
Default Value: STANdard
Example: To set correction type to flex:
:CORR:TYPE FLEX
[:SENSe]:CORRection[:STATe] ON|OFF|1|0
[:SENSe]:CORRection[:STATe]?
Cmd Description: Command sets correction status.
Query Description: Query returns correction status. Query returns 1 if RF cal is on.
Otherwise, it returns 0.
Param Range: ON|OFF|1|0
Default Value: OFF|0
Example: To turn calibration correction on:
:CORR:STAT ON
[:SENSe]:FREQuency:CABLe <cable-list-index>
[:SENSe]:FREQuency:CABLe?
Cmd Description: Command selects a cable with the index number from the sorted cable
list.
Query Description: Query returns the index number of the selected cable based on the
sorted cable list.
Param Type: Integer
Param Description: Index number from cable list
Default Value: 0
Example: To set the cable index to 10:
:FREQ:CABL 10
[:SENSe]:FREQuency:CABLe:CATalog:FULL?
Query Description: Query returns the entire cable list that is stored in memory.
Example: To get the cable list:
:FREQ:CABL:CAT:FULL?

Table of Contents

Other manuals for Anritsu Site Master S331L

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Site Master S331L and is the answer not in the manual?

Anritsu Site Master S331L Specifications

General IconGeneral
BrandAnritsu
ModelSite Master S331L
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals