EasyManuals Logo

Anritsu Site Master S331L Scpi Programming Manual

Anritsu Site Master S331L
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #63 background imageLoading...
Page #63 background image
Chapter 4 — Cable & Antenna Analyzer Mode Commands
S331L PM PN: 10580-00322 Rev. F 4-27
[:SENSe]:FREQuency:CABLe:NAMe <"cable-name">
[:SENSe]:FREQuency:CABLe:NAMe?
Cmd Description: Command sets the cable name.
Query Description: Query returns the cable name.
Param Type: String
Param Description: Cable name
Default Value: NONE
Example: To set the cable to "AVA5-50 7/8":
:FREQ:CABL:NAM "AVA5-50 7/8"
[:SENSe]:FREQuency:STARt <frequency>{ Hz|kHz|MHz|GHz}
[:SENSe]:FREQuency:STARt?
Cmd Description: Command sets the start frequency.
Query Description: Query returns the start frequency in Hertz.
Param Type: Float
Param Description: Start frequency
Default Value: Model-dependent
Default Units: Hz
Example: To set the start frequency to 2 GHz:
:FREQ:STAR 2 GHZ
[:SENSe]:FREQuency:STOP <frequency>{ Hz|kHz|MHz|GHz}
[:SENSe]:FREQuency:STOP?
Cmd Description: Command sets the stop frequency.
Query Description: Query returns the stop frequency in Hertz.
Param Type: Float
Param Description: Stop frequency
Default Value: Model-dependent
Default Units: Hz
Example: To set the stop frequency to 2 GHz:
:FREQ:STOP 2 GHZ

Table of Contents

Other manuals for Anritsu Site Master S331L

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Site Master S331L and is the answer not in the manual?

Anritsu Site Master S331L Specifications

General IconGeneral
BrandAnritsu
ModelSite Master S331L
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals