Chapter 3 — All Mode Commands
S331L PM PN: 10580-00322 Rev. F 3-9
[:SENSe]:BASe:NET:MODE:MANual:RESTart
Cmd Description: Command sets the Ethernet configuration to static (manual) and
restarts the instrument.
Example: To set the Ethernet configuration to static (manual) and trigger a
restart:
:BAS:NET:MODE:MAN:REST
[:SENSe]:BASe:REMote 0|1|2
Cmd Description: Command sets instrument into local lock-out for remote operation. Send
0 is to get out of remote mode and restore front-panel operation, 1 to set
remote mode with synchronous communication, and 2 to set remote
mode with asynchronous communication.
Param Type: Integer
Param Range: 1|0
Default Value: 0
Example: To set the instrument to remote synchronous mode and perform a local
lock-out:
:BAS:REM 1
[:SENSe]:BASe:SCRNshot:SCHeme INVerted|STANdard
[:SENSe]:BASe:SCRNshot:SCHeme?
Cmd Description: Command sets the screenshot setting scheme.
Query Description: Query returns the value that represents the current screenshot scheme.
Param Type: Character
Param Range: INVerted|STANdard
Default Value: INVerted
Example: To set the screen capture scheme to standard:
:BAS:SCRN:SCH STAN
[:SENSe]:BASE:TEMPerature?
Query Description: Query returns motherboard temperature in degree C. Same as
:SYSTem:MBTemperature?
Example: To query the motherboard temperature:
:BAS:TEMP?