EasyManuals Logo

Anritsu Site Master S331L Scpi Programming Manual

Anritsu Site Master S331L
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
Chapter 3 — All Mode Commands
S331L PM PN: 10580-00322 Rev. F 3-9
[:SENSe]:BASe:NET:MODE:MANual:RESTart
Cmd Description: Command sets the Ethernet configuration to static (manual) and
restarts the instrument.
Example: To set the Ethernet configuration to static (manual) and trigger a
restart:
:BAS:NET:MODE:MAN:REST
[:SENSe]:BASe:REMote 0|1|2
Cmd Description: Command sets instrument into local lock-out for remote operation. Send
0 is to get out of remote mode and restore front-panel operation, 1 to set
remote mode with synchronous communication, and 2 to set remote
mode with asynchronous communication.
Param Type: Integer
Param Range: 1|0
Default Value: 0
Example: To set the instrument to remote synchronous mode and perform a local
lock-out:
:BAS:REM 1
[:SENSe]:BASe:SCRNshot:SCHeme INVerted|STANdard
[:SENSe]:BASe:SCRNshot:SCHeme?
Cmd Description: Command sets the screenshot setting scheme.
Query Description: Query returns the value that represents the current screenshot scheme.
Param Type: Character
Param Range: INVerted|STANdard
Default Value: INVerted
Example: To set the screen capture scheme to standard:
:BAS:SCRN:SCH STAN
[:SENSe]:BASE:TEMPerature?
Query Description: Query returns motherboard temperature in degree C. Same as
:SYSTem:MBTemperature?
Example: To query the motherboard temperature:
:BAS:TEMP?

Table of Contents

Other manuals for Anritsu Site Master S331L

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Site Master S331L and is the answer not in the manual?

Anritsu Site Master S331L Specifications

General IconGeneral
BrandAnritsu
ModelSite Master S331L
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals