Chapter 4 — Cable & Antenna Analyzer Mode Commands
4-14 PN: 10580-00322 Rev. F S331L PM
:CALCulate:MARKer:PRESet
Cmd Description: Presets markers.
Example: To preset markers:
:CALC:MARK:PRES
:CALCulate:MATH:FUNCtion NONE|SUBTraction|ADDition|AVERage
:CALCulate:MATH:FUNCtion?
Cmd Description: Command sets the trace math function. NONE - no trace math applied;
SUBT - memory trace subtracted from active trace; ADD - memory trace
added to active trace; AVER - active and memory traces added and
divided by 2.
Query Description: Query returns the trace math function.
Param Range: NONE|SUBTraction|ADDition|AVERage
Default Value: NONE
Example: To set trace math function to addition:
:CALC:MATH:FUNC ADD
:CALCulate:MATH:MEMorize
Cmd Description: Copies active trace into the memory buffer.
Example: To copy active trace into the memory buffer:
:CALC:MATH:MEM
:CALCulate:TRANsform:CLAVerage?
Query Description: Query returns cable loss average when the measurement mode is ’Cable
Loss’.
Default Units: dB
Example: To query cable loss average:
:CALC:TRAN:CLAV?