Chapter 4 — Cable & Antenna Analyzer Mode Commands
S331L PM PN: 10580-00322 Rev. F 4-5
:CALCulate#:LIMit:LOWer[:STATe] OFF|ON|0|1
:CALCulate#:LIMit:LOWer[:STATe]?
Cmd Description: Command sets the low limit line state for the specified trace.
Query Description: Query returns the lower limit line state for the specified trace.
Param Range: OFF|ON|0|1
Default Value: OFF|0
Example: To set limit on for trace 1:
:CALC1:LIM:LOW ON
Notes: Replace ’#’ with the required trace number. Check User Guide for the
number of supported traces in each mode.
:CALCulate#:LIMit:UPPer:SEGMent:ADD
<StartX><StopX><StartY><StopY>
Cmd Description: Command adds a segment to the upper limit line.
Param Type: Float
Default Units: X units: Hz for Frequency domain, m or ft for Distance domain; Y units:
same as current active trace y-axis units
Example: To add an upper limit segment from 1MHz to 1GHz with value 20 dB for
trace 1:
:CALC1:LIM:UPP:SEGM:ADD 1e6,1e9,20,20
Notes: Replace ’#’ with the required trace number. Check User Guide for the
number of supported traces in each mode.
:CALCulate#:LIMit:UPPer:SEGMent:DELete
Cmd Description: Command deletes the active upper segment.
Example: To delete the active upper segment for trace 1
:CALC1:LIM:UPP:SEGM:DEL
Notes: Replace ’#’ with the required trace number. Check User Guide for the
number of supported traces in each mode.