EasyManuals Logo

Anritsu Spectrum Master MS2720T User Manual

Anritsu Spectrum Master MS2720T
Go to English
1035 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #474 background imageLoading...
Page #474 background image
Channel Scanner (Option 27) 4-6 Script Master Measurement Setup
Spectrum Analyzer MG PN: 10580-00349 Rev. H 4-5
4. Press the Repeat Scan Type to select the desired scan mode # Scans, go to step 4a, to
select Time, go to step 4b.
a. If # Scans is selected, press # of Repeats (List) to set the desired repetition of test
cycles of the Script Master Test File. Select the # of Repeats (List) define how
many channels to include in the custom list. This choice can be changed later if
needed.
b. If Time is selected, press Scan Duration to set the desired testing period. Time on
the submenu key turns red for editing. Pressing a number on the numeric keypad
lists the Time menu. Press the appropriate time unit.
5. If repeated set testing is desired, press the # of Repeats (Sets) submenu key and enter
the desired number of test cycles.
6. Press the Record to the On position to store test measurements.
7. Press the Start/Restart Test submenu key to begin testing.
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com

Table of Contents

Other manuals for Anritsu Spectrum Master MS2720T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Spectrum Master MS2720T and is the answer not in the manual?

Anritsu Spectrum Master MS2720T Specifications

General IconGeneral
BrandAnritsu
ModelSpectrum Master MS2720T
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals