2-3 Measurement Example Chapter 2 — E-Series Tracking Generator
2-2 PN: 10580-00339 Rev. C Tracking Generator MG
2-3 Measurement Example
This example demonstrates how to connect a Device Under Test (DUT) and make a
measurement. This example will characterize the frequency response of a filter in the GSM
frequency band.
1. Set the frequency range to cover the range of the DUT being tested. In this example, the
center frequency is set to 1.8 GHz with a 1 GHz span.
2. Set the Generator to Tracking Mode with an Output Power of 0.0 dBm. Refer to the
previous “Tracking Generator Overview” section for additional information.
3. Connect the DUT as shown in Figure 2-1 on page 2-2. For connector torque and
connector care information, refer to the User Guide for your instrument.
4. Toggle the Generator Output to On.
Note
Turning the Tracking Generator On sets the detection type (Amplitude > Detection)
to Sample and sets the sweep mode (Sweep > Sweep Mode) to No FFT. Turning
off the generator resets both parameters to their previous settings. Either parameter
may be changed by the user with the Tracking Generator on or by the instrument
when other measurements are also turned on.
Figure 2-1. Connection of Device Under Test
Band Pass
Filter
Power Charge
+/-
.
0
3
Sweep
2
Calibrate
1
Preset
6
Limit
5
Trace
4
Measure
9
Mode
8
System
7
File
Shift
Back
Enter
ESC
SpectrumMaster
MS2713E
GENERATOR/
RF OUT
ANALYZER/
RF IN
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com