7 Access Levels and Administrator Tasks
44 STERRAD VELOCITY Reader Users Guide
ASPACCESS Configuration
In the Utilities Menu touch ASP ACCESS Cong to see the ASP ACCESS
Conguration settings. Refer to ASP ACCESS Users Guide to generate a Pairing Code. In
ASP ACCESS Conguration ll in the ASP ACCESS Server, the IP Address, Username and
the Pairing Code. Touch Test to verify that the reader is connected to the ASP ACCESS
device. During the test the reader will display a busy screen with the words “One
Moment… Verifying Connection. If a connection is detected, the reader will display a
check mark in a green circle with the words “Connection Veried” for three seconds before
returning to the ASP ACCESS Conguration screen. If a connection is not detected, the
reader will display a yellow circle with an ”!” and the words “No Connection! Please verify
IP Address, Username and Pairing Codes are correct. for three seconds before returning
to the ASP ACCESS Conguration screen. After returning to ASP ACCESS Conguration,
verify that the settings are correct and test the connection again. If the test fails a second
time refer to the “Troubleshooting” section of this guide (Chapter 8). Touch Done to return
to the Utilities Menu.
Control Alerts
In the Utilities Menu touch Control Alerts. The Control Alerts utility is used to
determine the positive control BI request frequency. The reader is congured to request
a new positive control every 24 hours. The user congurable control alert default and
maximum time period is 24 hours. This time period can be changed so that the reader
will request a positive control more often.
The Chemical Indicator Color Change toggle switch allows the user to stop the
software from asking if the CI has changed color on a positive control BI. The chemical
indicator only changes color when exposed to hydrogen peroxide or other oxidizers, and
if stored properly the positive control BI’s chemical indicator should not have changed
color. Follow the applicable hospital or healthcare facility policy regarding control
chemical indicator color change documentation.