Troubleshooting 8
STERRAD VELOCITY Reader User’s Guide 47
Chapter 8.
Troubleshooting
User Error Messages
Problem/Error Message Possible Cause Suggested Solution
Please remove the
biological indicator and
wait for the reader to
complete its warm-up
sequence.
BI is placed in reader
before it has a
chance to complete
the warm-up
sequence.
Insertion sensor is
out of calibration.
Wait until the reader has
completed the warm-up
sequence before attempting to
insert BIs.
If BIs are not present then
perform an insertion sensor
calibration after the reader
has completed its warm
up sequence. Refer to the
“Calibrate Insertion Sensor”
section (Chapter 7) for more
information.
You haven’t nished
lling in the requested
information. Please
follow the directions on
the screen and ll in the
requested information.
User has not
conrmed BI result.
Enter the required information
and conrm the result. Failure
to do so will result in delay of
other BI results.
An incompatible
biological indicator
has been scanned. A
STERRAD VELOCITY
Biological Indicator
must be used in the
STERRAD VELOCITY
Reader.
Incompatible
biological indicator
used in the reader.
Use STERRAD VELOCITY
Biological Indicators.