EasyManua.ls Logo

Azbil MagneW FLEX+ - 1: Detector Position; Figure 2-1 Detector Placement; Figure 2-2 Straight Pipe Section on Upstream Side of Detector

Azbil MagneW FLEX+
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation Azbil Corporation
2-4 MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector
2.3.1: Detector position
Position the detector so that its inner detector passage is continuously lled with
the uid being measured. The following gure provides examples.
Pump
Bad
Air is easily trapped -
may not ll with uid
May not ll
with uid
Good
Good
Figure 2-1 Detector placement
CAUTION
The detector must be positioned as shown by the circled areas in the gure shown
above. If the pipe is not lled, output errors will occur.
When measuring high viscosity uid, connecting the pipe to a vertical detector is
recommended (in order to secure an axial symmetrical ow).
Install a straight pipe section on the upstream side of the detector. The following
gure shows the length of the straight pipe section (D = nominal bore diameter of
the detector).
Upstream
> 5D
T-joint
Gate valve (Completely open)
Diffuser with cone angle >15
(<15 Considered straight pipe section)
Concentrator
(Considered straight pipe section)
Any type of valve
Right angle joint
Detector
> 5D
Detector Detector
> 10D
Detector
> 5D
Detector
> 5D
Detector
> 5D
Figure 2-2 Straight pipe section on upstream side of detector
Although a pipe section is not necessary on the downstream side, secure a section
of at least 2D if eccentric ow appears.
Select an installation site where there is no major pulsation or vibration (away
from a pump).
Make sure that there is adequate space around the detector after installation to
perform maintenance.

Table of Contents

Related product manuals