Installation Azbil Corporation
2-6 MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector
2.4 : Installing a wafer detector
CAUTION
Before installing the detector, make sure any foreign matter is ushed from
the interior passage of the detector. Residual foreign matter could cause output
uctuations.
Do not touch the electrodes or allow oil or fat to come into contact with them to
avoid output uctuations.
Align the direction mark on the detector in the direction of the liquid ow.
Misalignment can result in a negative output.
The following gure shows the basic installation method for installing a wafer
detector.
Nuts (optional)
Through-bolts (optional)
Pipe
Gasket (Required
when using SUS
material grounding
rings. Otherwise,
a
gasket is
supplied.)
Centering nuts (supplied)
Figure 2-5 Installing a wafer detector