Safety instructions
properly inserted and primed. Discon-
nect it from the patient when chang-
ing infusion line in order to prevent
unintended dose administration.
Q
Only use approved infusion lines/
catheters for the intended medical use.
Q
Position the infusion line to the patient
so that it does not have any kinks.
Q
Ensure that installation in rooms
used for medical purposes is done in
accordance with the regulations (e.g.,
VDE 0100, VDE 0107 and/or IEC speci-
cations). Observe all country-specic
regulations and national deviations.
Q
Do not operate the device near inam-
mable anaesthetics.
Q
Always check the plausibility of the
values shown on the display.
Q
Ensure that there is additional patient
supervision (e.g. monitoring) if life sus-
taining drugs are administered.
Q
When administering highly-eective
drugs, have a second device ready for
the drug.
Q
Avoid mechanical eects on the
device. Ifthe device is moved while in
operation, the set
delivery rate may be
exceeded/not be reached.
Q
Irrespective of the soft limit, ensure
that the values set for the patients are
the medically correct values.
Q
When using the device near equipment
that can cause higher interference
emissions (e.g. electrosurgical devices,
magnetic resonance imaging units,
mobile telephones) keep the device
the recommended safe distance away
from such equipment.
status LEDs and the display during
theself-test.
Q
When xing the device to a box rail,
donot x the device near the rail
bracket.
Q
Fully charge the battery before the rst
use without an external power supply.
4.1.5
Stacking
Q
Stack a maximum of three devices on
top of one another.
Q
Do not stack in ambulances.
Q
When stacking, ensure that the device
is correctly and safely locked in. You
will hear an audible click sound when
the device is locked in.
WARNING! The use of this device adja-
cent to or stacked with other equip-
ment has to be avoided. Nevertheless,
if adjacent or stacked use is necessary,
the Infusomat compact
plus
P and the
other devices have to be observed to
verify normal operation in the con-
guration in which it will be used.
Note: A list of equipment with which the
Infusomat compact
plus
P has been tested
in a stacked or adjacent conguration and
with which stacked or adjacent use is per-
mitted can be found in section20.1.
4.1.6
Operation
Q
Stand in front of the device to operate
it. This ensures that you are able to
reach all control elements and that the
display is clearly visible.
Q
Establish a connection to the patient
only after the infusion line has been
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 11 15.10.2020 17:47:25