Alarms
Display
notication
Meaning
“Pressure
too high
There is an occlusion in
the system. The set level
was exceeded
Q
The pump automati-
cally implements a
bolus reduction
Q
Check that there are
no kinks in the tubing
and that it is undam-
aged and that there is
IV and lter patency
“KVO
nished
KVO time has elapsed
Q
Continue with therapy
or select new therapy
“No
battery in
the device”
It is not possible to use the
pump without a battery
Q
Ask a service technician
to insert a battery
“Air
b ubble/
accumu-
lated air”
Air in the system.
Q
Check the line for
small air bubbles
and, ifnecessary,
disconnect the patient
and prime again.
“Line
Calibration
Error”
The pump and infusion line
are not occluded.
Q
Close the roller clamp
and perform a line
replacement.
“Upstream
pressure
alarm”
The pressure on the
container-side is too low,
e.g. because the container
is empty.
9.2.2
Operating alarms
In the event of an operating alarm,
theinfusion is stopped. Anacoustic signal
sounds, the red LED ashes and a sta call
is activated.
Note: If an operating alarm is not acknowl-
edged within two minutes, another acoustic
signal sounds.
Display
notication
Meaning
“Target
volume
reached
Preselected volume has
been infused
Q
Continue with delivery
or start new therapy
Time
reached
Preselected time has
elapsed
Q
Continue with therapy
or select new therapy
“Battery
empty
The battery is discharged
Q
Connect device to
mains and/or have
battery replaced by a
service technician
The battery alarm will
sound for 3min. Then the
pump will automatically
turn o.
Note: If the battery has
deep discharged, the
notication may also say
“Battery warning”.