36
Alarms
Display
notication
Meaning
“Pressure
too high”
There is an occlusion in
the system. The set level
was exceeded
Q
The pump automati-
cally implements a
bolus reduction
Q
Check that there are
no kinks in the tubing
and that it is undam-
aged and that there is
IV and lter patency
“KVO
nished”
KVO time has elapsed
Q
Continue with therapy
or select new therapy
“No
battery in
the device”
It is not possible to use the
pump without a battery
Q
Ask a service technician
to insert a battery
“Air
b ubble/
accumu-
lated air”
Air in the system.
Q
Check the line for
small air bubbles
and, ifnecessary,
disconnect the patient
and prime again.
“Line
Calibration
Error”
The pump and infusion line
are not occluded.
Q
Close the roller clamp
and perform a line
replacement.
“Upstream
pressure
alarm”
The pressure on the
container-side is too low,
e.g. because the container
is empty.
9.2.2
Operating alarms
In the event of an operating alarm,
theinfusion is stopped. Anacoustic signal
sounds, the red LED ashes and a sta call
is activated.
Note: If an operating alarm is not acknowl-
edged within two minutes, another acoustic
signal sounds.
Display
notication
Meaning
“Target
volume
reached”
Preselected volume has
been infused
Q
Continue with delivery
or start new therapy
“Time
reached”
Preselected time has
elapsed
Q
Continue with therapy
or select new therapy
“Battery
empty”
The battery is discharged
Q
Connect device to
mains and/or have
battery replaced by a
service technician
The battery alarm will
sound for 3min. Then the
pump will automatically
turn o.
Note: If the battery has
deep discharged, the
notication may also say
“Battery warning”.
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 36 15.10.2020 17:47:29