Safety instructions
4.1.3
Transport and storage
Q
Devices stored at temperatures above
and below the dened operating
conditions range must be kept at room
temperature for at least one hour
before being powered on.
Q
The risk of transport damage cannot
be completely eliminated despite care-
ful packaging. Please check on delivery
that all items are present. Do not use
a damaged device. Contact the Service
department. Check that the device is
working correctly before its rst use.
This is even a legal requirement in sev-
eral countries. A corresponding form is
available from B. Braun.
4.1.4
Set-up and start-up
Q
For mobile use (patient transport within
the clinic and outside the clinic), ensure
secure mounting or positioning of the
device. Changes of position and strong
vibrations can cause minor changes
in the delivery characteristics and/or
unintentional bolus administration.
Q
Changing the height during a running
infusion may lead to ow rate varia-
tions.
Q
Ensure that the device is properly posi-
tioned and secured, and that it is level.
Q
Do not position the device above the
patient.
Q
Before switching on the device check
it, and the air sensor in particular, for
contamination, damage, missing parts
and functionality.
Q
Pay attention to audible and visible
alarms, the lighting up of the two
4 Safety instructions
Q
Read the safety instructions before
using the device and observe them.
4.1
Safe handling
4.1.1
General
Q
Make sure that the introductory
training on the device is given by
a B.Braun sales representative or
another authorised person.
Q
Any serious incident that has occurred
in relation to this product should be
reported to B. Braun and the compe-
tent authority of the country in which
the product is operated.
Q
If the device is dropped or subjected to
external forces: stop using the device
and have it tested by an authorised
service workshop.
Q
Protect the device against moisture.
Q
Keep the device clean.
Q
Close patient access in standby mode.
4.1.2
Software
Q
Consult the instructions for use fol-
lowing each software update to nd
out about the most recent changes to
the device and its accessories.
Q
Ensure that the software version on
the device corresponds to the version
these instructions for use refer to.
Q
Ensure that all devices used in a station
have the same software version
installed to avoid mistakes when using
dierently congured devices.