EasyManua.ls Logo

B. Braun Infusomat compact plus - 10.2 Battery operation and maintenance

B. Braun Infusomat compact plus
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
39
Cleaning and care
10.2
Battery operation
and maintenance
The device is equipped with a modern
lithium- ion battery that, atthe time of
delivery and after being fully charged,
guarantees an operating time of 6hours
at 25ml/h. For optimal treatment of the
battery, thedevice is equipped with protec-
tion against overcharge and deep depletion.
The battery is charged by the device during
mains operation.
In the event of a power cut or disconnection
from the mains, the pump automatically
switches to battery mode.
The battery status indicator in the display is
a trend display (low, medium, high).
10.2.1
Notes for optimal
battery operation
Battery life may vary due to
Q
Ambient temperature
Q
Varying loads
Therefore, please observe the following:
Q
Under normal temperature conditions,
abattery can be fully discharged and
recharged around 300 times before its
capacity decreases to around half of the
original nominal value.
Q
When the device is in mains operation
mode, the battery discharges slowly
and may be fully exhausted even if the
device is not in operation. In this case
the battery does not reach its original
capacity after one charge; it takes
several charging and discharging cycles
for the battery to achieve its original
capacity.
Group Active Substance
Peroxide Hydrogen Peroxide,
Peracetic Acid,
Mono peroxyphthalat-
hexahydrat
Aldehydes Glutaral,
Glyoxal,
Formaldehyde
Alkylamines N-(3-aminopropyl)-N-
Dodecylpropan-1,3-Diamin,
Cocospropylendiamin
If you have any questions about the use of
a particular disinfectant, please contact the
manufacturer of the respective disinfectant.
Note: The use of unapproved cleaners and
failure to follow the disinfection procedures
and the manufacturer’s recommended dilu-
tions can result in an instrument malfunc-
tion or product damage and could void the
warranty.
Q
No pointed objects should be used for
cleaning.
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 39 15.10.2020 17:47:30