34
Operation
Q
Remove the infusion line.
Q
Insert the new infusion line,
see section 8.2.
Q
Start the infusion, see section 8.4.
8.14
Ending the infusion
Q
Press the Start/Stop key to end the
infusion.
The green LED turns o.
Q
Disconnect the line from the patient
and close the roller clamp.
Q
Open the pump door by pulling the
door lever. Grip the door lever from
behind and pull it forward.
Q
Remove the infusion line.
Q
Close the pump door while simultane-
ously pulling the door lever.
8.15
Switching o the device
Q
Infusion ended
Note: The device cannot be switched o if
a disposable item is inserted. Instead it will
go into standby mode.
Q
Press the On/O key for approx.
1.5seconds.
The device switches o.
–
If the rate and time have been
entered, the remaining time is
adjusted.
Q
Time changed:
–
If only the rate has been entered,
the remaining volume is adjusted.
–
If only the volume has been
entered, the rate (or dose rate) is
adjusted.
–
If the rate and volume have been
entered, the remaining volume is
adjusted.
8.12
Resetting the therapy
The “Reset therapy” function is used to
delete all currently set therapy data. A new
therapy can be started.
Additionally, the volume already infused
can be reset.
Note: Reset therapy can only be selected if
the therapy has been stopped.
Q
Press the Menu key, select Reset
therapy and press OK to conrm.
Q
Press the up arrow key to reset the
therapy.
Q
Press the up arrow key to reset the
volume.
8.13
Changing the infusion line
Q
Press the Start/Stop key to stop the
infusion.
The green LED turns o.
Q
Disconnect the line from the patient
and close the roller clamp.
Q
Open the pump door by pulling the
door lever. Grip the door lever from
behind and pull it forward.
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 34 15.10.2020 17:47:29