44
Technical data
17 Technical data
Note: The data given, e.g. delivery accuracy, pressure alarm and alarm reaction times, apply
at room temperature and with water as the test material. Dierent media viscosities and
temperatures may lead to deviations.
Parameter Value
Type of device Volumetric infusion pump
Product classication According to Directive 93/42 EEC:
Q
IIb
According to EN 60601-1:
Q
Protection class II
Q
For debrillation-proof type CF applied part
Moisture protection IP34
Q
Protection against penetration by solid foreign bodies
with a diameter of more than 2.5mm
Q
Protection against splashes from all directions
Power supply
Q
100-240 V AC, 50-60 Hz, connection via power cable
orcompact
plus
station
Q
12 V DC 12V CP interface cable
Internal battery
Q
Battery life
Q
Recharging time
Lithium-ion battery
Q
Approx. 8h at 25ml/h
2.5h at 1000ml/h
Q
Approx. 3h
Power consumption 10 VA typ. / 40 VA max.
Under normal conditions (charged battery, room temperature,
25ml/h): <3.5W
Current consumption/
charging current
Q
Max. 0.4A
e
(typ. <0.1A
e
) at 100-240 V AC, 50-60 Hz
Q
Max. 1.5A (typ. <0.5 A) at 12VDC
Sta call Max. 24V / 0.5 A / 24 VA (VDE 0834)
EMC DIN EN 60601-1:2006 (IEC 60601-1:2005)
DIN EN 60601-1-2:2007 (IEC 60601-1-2:2007)
DIN EN 60601-2-24:2015 (IEC 60601-2-24:2012)
Time of operation 100% (continuous operation)