48
Electromagnetic compatibility
18.1
Electromagnetic interference emissions
The device is designed to be used in the electromagnetic environmental conditions described
below. Customers or users of the compact
plus
system or its components should ensure that
the system is being operated in such an environment.
Interference emission
measurements
Compliance Electromagnetic environment
guidelines
RF emission as per CISPR11 Group 1 /
Class B
(see Note 1/
Note 2 below)
The Infusomat compact
plus
P uses RF
energy for its internal functions only.
Assuch, its RF emissions rate is very low
and it is unlikely to interfere with nearby
electronic equipment.
Voltage uctuation/icker
emissions according to
IEC61000-3-3
Conforms The device is intended for use in all
establishments (including residential
areas and similar) directly connected to a
public power grid that also supplies build-
ings used for residential purposes.
Harmonic emissions
according to
IEC 61000-3-2
Not
applicable
Note1: The limits for interference emissions are measured with individual components as well
as with the maximum set-up (fully equipped compact
plus
system).
Note2: If Class A equipment is attached to the compact
plus
system, the compact
plus
system
will become Class A too. This equipment/system may cause radio interference or may disrupt
the operation of nearby equipment. It may be necessary to take mitigation measures, such
as re-orienting or relocating the compact
plus
system or shielding the location.
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 48 15.10.2020 17:47:31