Warranty
16 Start-up and
trumpet curves
16.1
Signicance in
clinical practice
Trumpet curves show the recorded maximum
and minimum deviations in ow rate com-
pared to the delivery rate per time interval.
In clinical practice, the trumpet curve makes
it easier for the treating doctor to decide
if the pump is suciently precise for the
administration of the desired drug.
Q
Reconcile drugs with short half lives,
inparticular, with the delivery accu-
racy in this period on the trumpet
curve.
The physiological eect of the drug can
be aected by the ow and the disposable
article.
Q
Ensure that the prescription is in line
with the start-up/trumpet curve and
the set ow rate.
15 Warranty
B. Braun provides 24 months warranty,
as from the date of delivery, for every
Infusomat® compact
plus
P. This covers repair
or replacement of parts damaged as a result
of design/manufacturing errors or material
defects.
Modications or repairs to the unit under-
taken by the user/operator or by third par-
ties invalidate the warranty.
The warranty does not cover the following:
Elimination of faults attributable to incor-
rect/unauthorized handling, or to normal
wear and tear.
Defective rechargeable battery packs can
be returned to B. Braun for further disposal.
WARNING! Do not modify this
equipment without authorization of
the manufacturer.