49
Electromagnetic compatibility
18.2
Electromagnetic immunity
The device is designed to be used in the electromagnetic environmental conditions described
below. Customers or users of the compact
plus
system or its components should ensure that
the system is being operated in such an environment.
Immunity tests Test level
EN 60601-1-2
EN 60601-2-24
Compliance
level
Electromagnetic
environment guidelines
Electrostatic
discharge (ESD)
according to
IEC 61000-4-2
Contact discharge
IEC 60601-1-2:
±6 kV
IEC 60601-2-24:
±8 kV
±6 KV
without
interference
±8 KV outage
with alarm
permitted
Floors should be wood,
concrete or ceramic tile.
If oors are covered with
synthetic material, the rela-
tive humidity should be at
least 30 %.
Air discharge
IEC 60601-1-2:
±8 kV
IEC 60601-2-24:
±15 kV
±8KV without
interference
±15KV outage
with alarm
permitted
Electrical fast
transient/ bursts
according to
IEC61000-4-4
for power supply
lines ±2 kV
±2 kV Mains power quality should
be that of a typical com-
mercial or hospital environ-
ment.
For input and output
lines ±1 kV
±1 kV
Surges according
to IEC 61000-4-5
Dierential mode
voltage ±1 kV
±1 kV Mains power quality should
be that of a typical com-
mercial or hospital environ-
ment.
Common mode
voltage ±2 kV
±2 kV