50
Electromagnetic compatibility
Immunity tests Test level
EN 60601-1-2
EN 60601-2-24
Compliance
level
Electromagnetic
environment guidelines
Voltage dips,
brief supply volt-
age interruptions
and uctuations
according to
IEC61000-4-11
< 5% UT ¹
for ½ periods
(>95% dip)
Complies
through
the use of
an internal
energy source
The supply voltage quality
should be the same as that
of a typical commercial or
hospital environment.
40% UT ¹
for 5 periods
(60% decline)
70 % UT ¹
for 25 periods
(30 % decline)
<5% UT ¹
for 5 s
(>95% dip)
Magnetic eld at
supply frequency
(50/60Hz)
according to
IEC61000-4-8
3 A/m 400 A/m Magnetic elds at the
supply frequency should
correspond to those typi-
cally found in commercial
and hospital environments.
Conducted RF
interference
according to
IEC61000-4-6
IEC 60601-1-2:
150kHz to 80MHz
3V
RMS
outside and
10V
RMS
in ISM
frequency bands
IEC 60601-2-24:
150 kHz to 80 MHz
10V
RMS
10V
RMS
150kHz to
80Mhz
in all
frequency
bands
Do not use portable radio
communications equipment
closer to the Infusomat
compact
plus
P (including
connection cables) than
the recommended safe
distance calculated using
the appropriate equation for
that frequency.
Recommended safe
distance: d=1.2 √P
1)
1)
UT is the AC mains voltage prior to test level application
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 50 15.10.2020 17:47:32