45
Technical data
Parameter Value
Acoustic alarm signal
sound pressure range
Nine available levels:
45dB(A) to 75dB(A)
Interfaces
Q
Cold connector for mains voltage
Q
Accessory port for interface cable 12V CP and sta call
Q
IrDA infrared for communication in the station and for
service
Operating conditions
Q
Temperature
Q
Relative air humidity
Q
Atm. pressure
Q
+18 °C … +35 °C / +64 °F ... +95 °F
Q
30% … 90% (without condensation)
0.54 … 1.06bar
Storage conditions
Q
Temperature
Q
Relative air humidity
Q
Atmospheric pressure
Q
–20 °C … +55 °C / -4 °F ... +131 °F
Q
20% … 90% (without condensation)
Q
0.5 … 1.06bar
Weight Approx. 2.0kg
Dimensions in mm
(W x H x D)
Approx. 229mm x 98mm x 225mm
(including pole clamp compact
plus
)
Safety check Every 2 years
Volume preselection 0.1ml - 9,999 ml in increments of 0.01ml
Time preselection 00:01 h - 99:59 h
Delivery accuracy typ. ± 5% according to IEC 60601-2-24
Note: valid for 50cm water column
Occlusion alarm pressure 9 levels up to 1 bar
Alarm in the case of
incorrect dose
In the event of an incorrect dose of 1ml due to pump
malfunction (electronics, software), the pump will automati-
cally switch o.
Delivery rate increments 0.1ml/h ... 1000ml/h in increments of 0.01ml/h
Delivery accuracy for
bolus administration
typ. ± 5% for bolus volumes > 5ml
Max. bolus volume after
bolus reduction
≤ 0.2ml at rates ≥ 1ml/h